Aquaplore 3S

Unite as One    All in one

Aquaplore 3S is designed with a flexible modular architecture that easily meets the changing demands of influent quality and high volume ultrapure water, and provides consistent ultrapure water quality compliance and affordability for multi-user scenarios, avoiding additional expenses when demands change and allowing one system to easily adjust to meet your varying needs. No other system can be so perfect to meet the various needs of your team.

With Aquapore 3S you can ...

  • Track individual user usage history with password login
  • Reduce the cost of ultrapure water with an integrated purification unit with high throughput
  • Adapt to changes in raw water quality and usage requirements without purchasing new equipment using a modular architecture
  • Mobile water collection design for unlimited distance in multi-laboratory environments
 (Learn more)

Single/Multi-User Scenario

Easy account management to meet your compliance requirements, password-protected access mechanism, all user records are stored independently in the system. Switch freely between single/multi-user modes according to your usage patterns, allowing multiple departments and projects to share a single device and share the cost of supplies and maintenance.

Free combination of multifunctional standardized modules

Integrated structural design provides more capacity in the smallest size than ever before, effectively reducing replacement frequency and water costs
The standardized honeycomb modules with different purification functions can be freely combined according to different usage requirements, easily adapting to all your applications and demand adjustments without the need to purchase additional equipment.

Unique DP-RO Reverse Osmosis Technology

The patented DP- RO deep desalination technology can enhance the total desalination rate of the system to more than 99.5%, reduce the impact of changes in raw water quality, and steadily produce pure water in accordance with ISO standards when the conductity of raw water does not exceed 1000μS/cm, while significantly improving the durability of the system, extending the life of the ultrapure filter, and reducing the production cost of ultrapure water, solving the It also reduces the cost of producing ultrapure water and solves the high purchase and maintenance costs associated with continuous electrolytic desalination technology. Compared with electrolytic deionization (EDI) technology, DP-RO can quickly purify multiple batches of compliant pure water without a large buffer tank, and is ready for use.

Mobile closed type ultra-pure water tank

It solves the problem of piping contamination and distance limitation of traditional split water collection device, and saves the cost of expensive water collection device, especially suitable for multiple laboratories to share one ultra-pure water equipment. Patented respiratory filtration device, eliminating bacteria, particles and laboratory volatile gas contamination. The patented ergonomic handle makes it easy to place the tank where you need it.
Water quality
OR CE BE
Type I Water Resistivity (25°C ) 18.2MΩ.cm 18.2MΩ.cm 18.2MΩ.cm
Pure Water Resistivity (25°C) 10-15MΩ.cm 10-15MΩ.cm 10-15MΩ.cm
Total organic carbon1 (TOC) (treated by DP-RO) ≤2ppb ≤2ppb
Bacteria2 <0.001cfu/ml <0.001cfu/ml
Bacterial endotoxin2 <0.001EU/ml
Particles(>0.2μm) <1/ml <1/ml <1/ml
Inorganic ion removal rate ≥99.5% ≥99.5% ≥99.5%
Heavy metal ions ≤0.1ppb ≤0.1ppb ≤0.1ppb
Silicon removal rate >99% >99% >99%
RNases2 <1pg/ml
DNases2 <4pg/ml
1. Under the condition of feed water TOC < 50ppb and proper operation conditions.
2. With UV device and ultra-precision filtration module, under regular cleaning and maintenance.
Water production volume specifications
Pure water production volume (25°C)     5/10/20L/h
Quantitative water wi thdrawal range     10ml-99L, the value can be set arbitrarily, the accuracy is up to 土2ml within 500ml, 土1% above 500ml
Water withdrawal flow rate     ≥2L/min
Water utilization rate     75-80%
Water intake requirements
Water Source     tap water
Flow rate     ≥100L/h
Conductivity(25°C)     <2000μS/cm
Pressure     1-5bar
Temperature     5-35°C
Free chlorine     <1ppm
PH value     6.5-8.5
Total hardness(by CaCO3)     <450ppm
Electricity requirement
Electricity input     100-240VAC,50-60Hz
System voltage     24VDC
System power     221W(max)
Noise level     <50dB
Size and weight
Dimension(H×W×D)     550×420×550mm,The depth of host top shall not exceed 660mm
Packing weight     51Kg