Aquaplore 3

Easy access to compliant laboratory water

Aquplore3 is a system designed specifically for users who use smaller volumes but have strict compliance and quality requirements. It automatically monitors and traces your water production process and maintenance to ensure that your ultrapure water is of consistent quality and compliance. A variety of clever and intelligent designs help you quickly complete otherwise complex calibration and verification tasks, saving you unnecessary labor and time and allowing you to focus on more important research. Now you don't have to compromise your management efforts due to resource constraints, and you can easily have quality laboratory ultrapure water.

With Aquaplore 3 you can...

  • Perform system validation, calibration and complete system disinfection directly on-site. With the unique Process Integration Module (P.I.M.), system maintenance downtime can be significantly reduced
  • Complete monitoring and management of key parameters in the purification process to ensure reliable water quality.
  • Customize your maintenance cycle with an intelligent consumables management system to make your consumables replacement more economical and reasonable.
  • Choose the most suitable water intake method for your needs from a variety of intake modes.
 ( Learn more)

Water quality
OR CE BE
Type I Water Resistivity (25°C ) 18.2MΩ.cm 18.2MΩ.cm 18.2MΩ.cm
Grade III Water Conductivity1(25°C) 1-5μS/cm 1-5μS/cm 1-5μS/cm
Total organic carbon2 (TOC)
(treated by DP-RO)
≤2ppb ≤2ppb
Bacteria3 <0.001cfu/ml <0.001cfu/ml
Bacterial endotoxin3 <0.001EU/ml
Particles(>0.2μm) <1/ml <1/ml <1/ml
Inorganic ion removal rate ≥99.5% ≥99.5% ≥99.5%
Heavy metal ions ≤0.1ppb ≤0.1ppb ≤0.1ppb
Silicon removal rate >99% >99% >99%
RNases3 <1pg/ml
DNases3 <4pg/ml
1. In general municipal water supply with conductivity < 1000μS/cm condition.
2. Under the condition of feed water TOC < 50ppb and proper operation conditions.
3. With UV device and ultra- precision filtration module, under regular cleaning and maintenance.
Water production volume specifications
Secondary reverse osmosis water production volume (25°C)     5L/h、10L/h、20L/h
Quantitative water wi thdrawal range     10ml-99L, the value can be set arbitrarily, the accuracy is up to 土2ml within 500ml, 土1% above 500ml
Water withdrawal flow rate     ≥2L/min
Water utilization rate     75-80%
Water intake requirements
Water Source     tap water
Flow rate     ≥100L/h
Conductivity(25°C)     <2000μS/cm
Pressure     1-5bar
Temperature     5-35°C
Free chlorine     <1ppm
PH value     6.5-8.5
Total hardness(by CaCO3)     <450ppm
Electricity requirement
Electricity input     100-240VAC,50-60Hz
System voltage     24VDC
System power     221W(max)
Noise level     <50dB
Size and weight
Dimension(H×W×D)     647×410×470mm
Packing weight     40Kg