Aquaplore DURO

The easiest tertiary water supply solution

Aquaplore DURO uses tap water as the water source to produce reliable pure water at a rate of up to 20L/h in accordance with ISO 3696 Class III water specifications. The unique DP- RO technology effectively increases water utilization and reduces purification costs, while avoiding the effects of raw water quality variations, resulting in more consistent and reliable water purity. The highly efficient purification technology can quickly meet the large volume of water needed for general applications, and is especially suitable for general education or for use with various automatic instruments.

With Aquapore DURO you can ...

  • Reduce the cost of non-critical applications with unique DP-RO technology to produce compliant tertiary water
  • Achieve unlimited distance access to water in multi-lab environments with mobile access design
  • Choose from convenient rinse modes to quickly clean a variety of labware
  • Use password login to track individual user's usage history.
  • Application: Ware cleaning, ultra pure water supply, high pressure sterilizer, heating pot, constant temperature water bath, argon generator, ultrasonic cleaner, hydroponics.
(Learn more)

Water quality
OR CE BE
Grade III Water Conductivity(25°C) 1-5μS/cm 1-5μS/cm 1-5μS/cm
Total organic carbon2 (TOC)
(treated by DP-RO)
<3ppb <3ppb
Baecteeria3 <0.1cfu/ml <0.001cfu/ml
Baecteerial endotoxin3 <0.001EU/ml
Particles(>0.2μm) <1/ml <1/ml <1/ml
Inorganic ion removal rate ≥99.5% ≥99.5% ≥99.5%
Silicon removal rate >99% >99% >99%
1. In general municipal water supply with conductivity < 1000μS/cm condition.
2. Under the condition of feed water TOC < 50ppb and proper operation conditions.
3. With UV device and ultra- precision filtration module, under regular cleaning and maintenance.
Water production volume specifications
Secondary reverse osmosis water production volume (25°C)     50L/h、10L/h、20L/h
Quantitative water withdrawal range     10ml-99L,the value can be set arbitrarily, the accuracy is up to 土2ml within 500ml,土1% above 500ml
Water withdrawal flow rate     ≤2L/min
Water utilizaation rate     60-80%
Water intake requirements
Water Source     tap water
Flow rate     ≥100L/h
Conductivity(25°C)     <2000μS/cm
Pressure     1-5bar
Temperature     5-35°C
Free chlorine     <1ppm
PH value     6.5-8.5
Total hardness(by CaCO3)     <450ppm
Electricity requirement
Electricity input     100-240VAC,50-60Hz
System voltage     24VDC
System power     221W(max)
Noise level     <50dB
Size and weight
Dimension(H×W×D)     647×410×470mm
Packing weight     40Kg